9170 Series
Metrology Well Calibrators
|
Click here for a virtual
demo of a Metrology Well.
Requires Macromedia Flash player.
|
 |
Download the complete data sheet
with ordering information in
Acrobat (.pdf) format.
Right click and
Save Target As...
|
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the
world
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from 45 °C to 700 °C
| |
9170 |
9171 |
9172 |
9173 |
Range (at 23 °C ambient) |
45 °C to 140 °C
(49 °F to 284 °F) |
30 °C to 155 °C
(22 °F to 311 °F) |
35 °C to 425 °C
(95 °F to 797 °F) |
50 °C to 700 °C
(122 °F to 1292 °F) |
Display Accuracy |
±0.1 °C full range |
±0.1 °C at 100 °C
±0.15 °C at 225 °C
±0.2 °C at 425 °C |
±0.2 °C at 425 °C
±0.25 °C at 660 °C |
Stability |
±0.005 °C full range |
±0.005 °C to 100 °C
±0.008 °C to 225 °C
±0.01 °C to 425 °C |
±0.005 °C to 100 °C
±0.01 °C to 425 °C
±0.03 °C to 700 °C |
Axial Uniformity (60 mm) |
±0.1 °C at 45 °C
±0.04 °C at 35 °C
±0.02 °C at 0 °C
±0.07 °C at 140 °C |
±0.025 °C at 30 °C
±0.02 °C at 0 °C
±0.07 °C at 155 °C |
±0.05 °C at 100 °C
±0.1 °C at 225 °C
±0.2 °C at 425 °C |
±0.1 °C at 100 °C
±0.25 °C at 425 °C
±0.4 °C at 700 °C |
Radial Uniformity |
±0.01 °C full range |
±0.01 °C at 100 °C
±0.02 °C at 225 °C
±0.025 °C at 425 °C |
±0.01 °C at 100 °C
±0.025 °C at 425 °C
±0.04 °C at 700 °C |
Loading Effect (with a 6.35 mm reference
probe and three 6.35 mm probes) |
±0.02 °C at 45 °C
±0.005 °C at 35 °C
±0.01 °C at 140 °C
|
±0.005 °C at 30 °C
±0.005 °C at 0 °C
±0.01 °C at 155 °C |
±0.01 °C full range |
±0.02 °C at 425 °C
±0.04 °C at 700 °C |
Hysteresis |
0.025 °C |
0.04 °C |
0.07 °C |
Well Depth |
160 mm (6.3 in) |
203 mm (8 in) |
Resolution |
0.001 °C |
Display |
LCD, °C or °F, user-selectable |
Key Pad |
Ten key with decimal and +/- button.
Function keys, menu key, and °C / °F key. |
Cooling Time |
44 min: 23 °C to 45 °C
19 min: 23 °C to 30 °C
19 min: 140 °C to 23 °C |
30 min: 23 °C to 30 °C
25 min: 155 °C to 23 °C |
220 min: 425 °C to 35 °C
100 min: 425 °C to 100 °C |
235 min: 700 °C to 50 °C
153 min: 700 °C to 100 °C |
Heating Time |
32 min: 23 °C to 140 °C
45 min: 45 °C to 140 °C |
44 min: 23 °C to 155 °C
56 min: 30 °C to 155 °C |
27 min: 35 °C to 425 °C |
46 min: 50 °C to 700 °C |
Size (height x width x depth) |
366 x 203 x 323 mm (14.4 x 8 x 12.7 in) |
Weight |
15 kg (33 lb) |
15 kg (33 lb) |
13.2 kg (29 lb) |
15 kg (33 lb) |
Power |
115 VAC (±10%), 6.3 A, or
230 VAC (±10%), 3.15 A |
115 VAC (±10%), 10 A, or 230 VAC (±10%),
5 A |
Traceable Calibration (NIST) |
Data at 45 °C, 0 °C, 50 °C,
100 °C, and 140 °C |
Data at 30 °C, 0 °C, 50 °C,
100 °C, and 155 °C |
Data at 100 °C, 150 °C, 250 °C,
350 °C, and 425 °C |
Data at 100 °C, 200 °C, 350 °C,
500 °C, and 660 °C |
Calibrated to 660 °C;
reference thermometer recommended at higher temperatures. |
Specifications
|
Built-in
Reference Input |
Temperature Range |
200 °C to 962 °C
(328 °F to 1764 °F) |
Resistance Range |
0 W to 400 W, auto-ranging |
Characterizations |
ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11
Callendar-Van Dusen (CVD): R0, a, b, d |
Resistance Accuracy |
0 W to 20 W: 0.0005 W
20 W to 400 W: 25 ppm |
Temperature Accuracy (does not include
probe uncertainty) |
10 W PRTs:
±0.013 °C at 0 °C
±0.014 °C at 155 °C
±0.019 °C at 425 °C
±0.028 °C at 700 °C |
25 W and 100 W PRTs:
±0.007 °C at 100 °C
±0.006 °C at 0 °C
±0.011 °C at 155 °C
±0.013 °C at 225 °C
±0.019 °C at 425 °C
±0.027 °C at 661 °C |
Resistance Resolution |
0 W to 20 W: 0.0001 W
20 W to 400 W: 0.001 W |
Measurement Period |
1 second |
Probe Connection |
4-wire with shield, 5-pin DIN connector |
Calibration |
NVLAP accredited (built-in reference input
only), NIST-traceable calibration provided |
Every once in a while, a new product comes around that changes the rules. It happened
when we introduced handheld dry-wells. It happened when we introduced Micro-Baths. Now
weve combined bath-level performance with dry-well functionality and legitimate
reference thermometry to create Metrology Wells.
With groundbreaking new proprietary electronics from Flukes Hart Scientific
Division (patents pending), Metrology Wells let you bring lab-quality performance into
whatever field environment you might work in. New analog and digital control techniques
provide stability as good as ±0.005 °C. And with dual-zone control, axial (or
vertical) uniformity is as good as ±0.02 °C over a 60 mm (2.36 in)
zone. (Thats 60 mm!) Such performance doesnt exist anywhere else outside of
fluid baths.
In short, there are six critical components of performance in an industrial heat source
(which the European metrology community explains, for example, in the document EA-10/13):
calibrated display accuracy, stability, axial (vertical) uniformity, radial (well-to-well)
uniformity, impact from loading, and hysteresis. We added a seventh in the form of a
legitimate reference thermometer input and created an entirely new product category:
Metrology Wells.
(By the way, Metrology Wells are the only products on the market supported by published
specifications addressing every performance category in the EA-10/13. Our specs
arent just hopes or guidelines. They apply to every Metrology Well we sell.)
Display accuracy
Dry-wells are typically calibrated by inserting a calibrated PRT into one of the wells
and making adjustments to the calibrators internal control sensor based on the
readings from the PRT. This has limited value because the unique characteristics of the
reference PRT, which essentially become calibrated into the calibrator, are
often quite different from the thermometers tested by the calibrator. This is complicated
by the presence of significant thermal gradients in the block and inadequate sensor
immersion into blocks that are simply too short.
Metrology Wells are different. Temperature gradients, loading effects, and hysteresis
have been minimized to make the calibration of the display much more meaningful. We use
only traceable, accredited PRTs to calibrate Metrology Wells and our proprietary
electronics consistently demonstrate repeatable accuracy more than ten times better than
our specs, which range from ±0.1 °C at the most commonly used temperatures to
±0.25 °C at 661 °C.
An application note is available to help better understand the uncertainties mentioned
above. Click here (right click and Save Target As...), Understanding the
uncertainties associated with the use of Metrology Wells to download the
application note in Adobe Acrobat (.pdf) format.
Built-In Reference Thermometry!
Flukes Hart Scientific Division has been making the worlds best
thermometer readout devices for quite some time. Our Super-Thermometer,
Black Stack, and Tweener
thermometers are well-known everywhere. Now were making our proprietary Tweener
measurement circuitry available directly in a heat source our new Metrology Wells.
This optionally built-in input accepts 100-, 25-, and 10-ohm PRTs. It reads
thermometer probes accurately from ±0.006 °C at 0 °C to ±0.027 °C at
661 °C, not including errors from the probe. It is compatible with every PRT sold by
Hart and connects to Metrology Wells via a 5-pin DIN connector.
Two things dramatically differentiate the Tweener circuit from the measurement
electronics built into many dry-wells. First, it accepts unique ITS-90 characterization
coefficients from reference thermometers, which allow you to take full advantage of the
accuracies of those thermometers. Second, it comes with a traceable, accredited
calibration, providing you full confidence in the integrity of its measurements.
Nothing beats a Hart Metrology Well for industrial thermal performance. And nothing
beats a Tweener measurement for built-in reference thermometry. |
For even better accuracy, Metrology Wells may be ordered with built-in electronics for
reading external PRTs with ITS-90 characterizations. (See sidebar, Built in Reference
Thermometry)
Stability
Heat sources from Hart have long been known as the most stable heat sources in the
world. It only gets better with Metrology Wells. Both low-temperature units (Models 9170
and 9171) are stable to ±0.005 °C over their full range. Even the 700 °C unit
(Model 9173) achieves stability of ±0.03 °C. Better stability can only be found in
fluid baths and primary fixed-point devices. The off-the-shelf controllers
used by most dry-well manufacturers simply cant provide this level of performance.
Axial uniformity
The EA-10/13 document suggests that dry-wells should include a zone of maximum
temperature homogeneity, which extends for 40 mm (1.54 in), usually at the bottom of a
well. Metrology Wells, however, combine our unique electronics with dual-zone control and
more well depth than is found in dry-wells to provide homogeneous zones over 60 mm (2.36
in). Vertical gradients in these zones range from ±0.02 °C at 0 °C to
±0.4 °C at 700 °C.
Whats more, Metrology Wells actually have these specifications published for each
unit, and we stand by them.
Radial uniformity
Radial uniformity is the difference in temperature between one well and another well.
For poorly designed heat sources, or when large-diameter probes are used, these
differences can be very large. For Metrology Wells, we define our specification as the
largest temperature difference between the vertically homogeneous zones of any two wells
that are each 6.4 mm (0.25 in) in diameter or smaller. The cold units (9170 and 9171)
provide radial uniformity of ±0.01 °C and the hot units (9172 and 9173) range from
±0.01 °C to ±0.04 °C (at 700 °C).
Loading
Loading is defined as the change in temperature sensed by a reference thermometer
inserted into the bottom of a well after the rest of the wells are filled with
thermometers, too.
For Metrology Wells, loading effects are minimized for the same reasons that axial
gradients are minimized. We use deeper wells than found in dry-wells. And we utilize
proprietary dual-zone controls. Loading effects are as minimal as ±0.005 °C in the
cold units.
Hysteresis
Thermal hysteresis exists far more in internal control sensors than in good-quality
reference PRTs. It is evidenced by the difference in two external measurements of the same
set-point temperature when that temperature is approached from two different directions
(hotter or colder) and is usually largest at the midpoint of a heat sources
temperature range. It exists because control sensors are typically designed for ruggedness
and do not have the strain free design characteristics of SPRTs, or even most
PRTs. For Metrology Wells, hysteresis effects range from 0.025 °C to 0.07 °C.
Immersion depth
Immersion depth matters. Not only does it help minimize axial gradient and loading
effects, it helps address the unique immersion characteristics of each thermometer tested
in the heat source. Those characteristics include the location and size of the actual
sensor within the probe, the width and thermal mass of the probe, and the lead wires
used to connect the sensor to the outside world. Metrology Wells feature well depths of
203 mm (8 in) in the Models 9171, 9172, and 9173. The Model 9170 is 160 mm (6.3 in) deep
to facilitate temperature of 45 °C.
Other great features
A large LCD display, numeric keypad,
and on-screen menus make use of Metrology Wells simple and intuitive. The display shows
the block temperature, built-in reference thermometer temperature, cutout temperature,
stability criteria, and ramp rate. The user interface can be configured to display in
English, French, or Chinese.
All four models come with an RS-232 serial interface and the Model 9930, Interface-it
software. All are also compatible with Model 9938 MET/TEMP II software for completely
automated calibrations of RTDs, thermocouples, and thermistors (Metrology Wells with
built-in reference input options will be compatible with MET/TEMP II in early 2006).
Even without a PC, Metrology Wells have four different preprogrammed calibration tasks
that allow up to eight temperature set points with ramp and soak times between
each. There is an automated switch test protocol that zeros in on the
dead-band for thermal switches. And a dedicated °C/°F button allows for easy
switching of temperature units. 
Any of six standard inserts may be ordered with each unit, accommodating a variety of
metric- and imperial-sized probe diameters. (See inset at right. Download
the complete data sheet to view details.) And Metrology Wells are small enough and
light enough to go anywhere.
9170
The Model 9170 achieves the lowest temperatures of the series, reaching
45 °C in normal room conditions. The 9170 is stable to ±0.005 °C over
its full temperature range (up to 140 °C) and has 160 mm (6.3 in) of immersion
depth. With axial uniformity of ±0.02 °C and radial uniformity of ±0.01 °C,
this model delivers exceptional uncertainty budgets and is perfect for a variety of
pharmaceutical and other applications.
9171
If you need more depth, the Model 9171 provides 203 mm (8 in) of immersion over
temperatures from 30 °C all the way to 155 °C with full-range stability
of ±0.005 °C. Just like the 9170, this dry-well has exceptional axial and radial
uniformity. The display of the 9171 is calibrated to an accuracy of ±0.1 °C over
its full range.
9172
The Model 9172 provides temperatures from 35 °C to 425 °C with a calibrated
display accurate to ±0.2 °C at 425 °C. In addition to exceptional accuracy,
the 9172 is stable from ±0.005 °C to ±0.01 °C, depending on temperature.
With 203 mm (8 in) of immersion, the 9172 significantly reduces stem conduction errors at
high temperatures.
9173
For work between 50 °C and 700 °C, the Model 9173 provides unmatched
performance. The 9173 has a display accuracy of ±0.25 °C at 700 °C and an
immersion depth of 203 mm (8 in). Stability and uniformity performance of this unit are
enough to dramatically reduce uncertainty budgets for calibrations of thermometers at high
temperatures.
Of course, theres still a place in the world for dry-wells or dry
block calibrators. In fact, Hart makes and will continue to make some of the best
performing, portable, fast dry-wells in the world. Theres still nothing better for a
quick test of industrial temperature sensor performance.
We just cant resist the urge, though, to keep coming up with breakthrough product
designs that can dramatically impact the ways people work and the results they see. For
the absolute best performance in a portable temperature source, Metrology Wells raise the
standard to an entirely new level.
For more information or for a price quote use our feedback form.
|
|